Visually directed high-intensity focused ultrasound for organ-confined prostate cancer: a proposed standard for the conduct of therapy

27 July, 2006

Prostate cancer is the most common cancer in men and the second leading cause of death from malignancy in the UK1. The mainstay of treatment remains radical surgery or radiation therapy, but several minimally invasive treatments are now under evaluation that might prove to be of equivalent oncological effectiveness in the long term2. Transrectal high-intensity focused ultrasound (HIFU) is one such treatment that has been used on an experimental and clinical basis as noninvasive therapy for clinically localized prostate cancer since the 1990s3.

HIFU relies on the physical properties of ultrasound energy. For therapeutic purposes it is focused by either an acoustic lens, bowl- shaped transducer or electronic phased array.

As ultrasound propagates through tissue, zones of high and low pressure are created. When the energy density (also known as focal intensity, measured in W/cm2) at the focus is sufficiently high (during the high-pressure phase), tissue damage can occur as a result of thermal coagulation necrosis and/or acoustic cavitation. The volume of a HIFU-generated lesion at the focal point is small (typically 10 mm long by 1–2 mm wide, in a cigar shape orientated along the long axis of the beam). If the intention is to ablate a given volume of tissue, individual lesions are placed next to each other to provide a continuous zone of necrosis.

It was shown experimentally that when mammalian tissue at the focus of a HIFU beam is raised to >60 °C for 3 s, all of the cells in that volume are rendered nonviable4. Thethreshold for achieving this is thought to be relatively constant among subjects5. Accordingly, algorithms were developed assuming certain tissue-related properties, tissue homogeneity and fixed ultrasound absorption coefficients that aim to produce thermal ablation using predefined power/time combinations at given tissue depths. In reality, the HIFU beam propagates through tissue and tissue interfaces that are characterized by natural variability, e.g. prostates vary among persons in size and in the ratio of stroma to epithelium. This will effect absorption coefficients and attenuation. Moreover, the presence of disease (cancer or no cancer) and the androgenic status of a patient are likely to add to this variability. These facts make it unlikely that an algorithm-based method of treatment will be the most likely to achieve the desired effects in most patients.

It follows therefore that some method is required for adjusting the energy to suit the unique characteristics of the prostate being treated. It is generally accepted that real-time imaging is a desirable attribute for any new minimally invasive therapy6, but there is debate about the best method to use. B-mode ultrasonography (US) is the only method in clinical use for monitoring HIFU therapy of the prostate, and this relies on detecting hyperechoic grey-scale changes within the treatment field. These changes are the result of both acoustic cavitation and tissue water vaporization, the latter occurring at boiling point. Grey-scale changes seen on B-mode US were correlated with histological changes within treated tissue during extracorporeal7 and transrectal therapy8, and their formation postulated for use in the control of prostate ablation9, but they have not been formally categorized to aid the clinician in conducting the therapy.

We describe our early experience of HIFU therapy using two distinct approaches to treatment. The first regimen was based on an estimated energy exposure, the algorithm- based approach; the second actively sought to generate grey-scale changes and to use these to guide energy exposure to the prostate. We described this type of treatment as ‘visually directed’. In addition to describing the outcomes of care associated with these two approaches, we propose a standardized nomenclature for the changes seen on B- mode US imaging during HIFU therapy for prostate cancer.

Patients and Methods

Between November 2004 and October 2005, 61 men were treated using the Sonablate500® (Focus Surgery, IN, USA) which consists of a power generator, water-cooling system (the ‘Sonachill®’), a treatment probe and a probe-positioning system (Fig. 1). The probe has two curved rectangular piezoceramic transducers with a driving frequency of 4 MHz and focal lengths of 30 and 40 mm, respectively. During treatment, these can be driven at low energy to provide real-time diagnostic US imaging or at high energy for therapeutic ablation (in situ intensity 1300–2200 W/cm2). The probe is covered by a condom through which cold (17–18 °C) degassed water circulates pumped by the Sonachill.

Thirty-four of the 61 men treated were included in this report (Fig. 2). All had prostate cancer stage =T2 (N0,M0), a PSA level of <15 ng/mL and prostate gland volumes of 1 cm diameter, as visualized by a previous TRUS. Written informed consent was obtained before treatment in all cases, and all men were followed-up for =3 months. It was necessary to exclude from the analysis men who had previously had hormone therapy, as this would confound the PSA nadir recorded after therapy.

Men were prepared before the procedure with two phosphate enemas to empty the rectum; an oral bowel preparation was used in some cases. Treatment was under general anaesthesia in all cases to reduce patient movement and discomfort. Men were placed in the lithotomy position, and the anal sphincter gently dilated. The treatment probe was introduced with a covering of ultrasound gel to couple it to the rectal mucosa, and then held in position by an articulated arm attached to the theatre table. A 16 F Foley urethral catheter was inserted under sterile technique, and a 10 mL balloon inflated to allow accurate visualization of the bladder neck and median sagittal plane.

Axial and sagittal US images were taken through the prostate using the transducer in the diagnostic mode. Treatment planning used proprietary software, which allows the prostate to be divided into ‘blocks’: anterior, middle and posterior, on both right and left sides. The software directs the transducer to move automatically so that the acoustic focus is moved sequentially through each point in the block. Each acoustic pulse ablates a volume of 3 × 3 × 10 mm, by heating the tissue to 80–98 °C almost instantaneously10, and individual lesions overlap slightly to ‘paint out’ the entire volume, using a combination of 3-s exposures (‘on’) time and 6-s pauses (‘off’) time, during which the gland was visualized with real-time US. The 4-cm focal length probe was used to treat anterior and middle blocks, and the 3-cm probe used to treat the posterior block.

The software is semi-automated, with the amount of energy applied to the prostate remaining under the control of the user. As a result, it is possible to treat the prostate in several ways. For instance, one approach uses pre-set energy exposure levels, the intensity of which depends on the part of the prostate that was being treated, and whether the treatment is a primary or salvage (after radiation) case. To a large extent, these energy exposure levels are derived from animal experiments11 or as a result of outcome monitoring in case series12. This might be termed an algorithm-based approach. Clinical series using this technique showed that the mean PSA nadirs achievable after treatment were ˜1.4 ng/mL13. These results are similar to those achieved by other transrectal HIFU devices that rely on the upper power limit being set without user control14.

An alternative method of managing energy exposure might involve abandoning any preset criteria to permit the maximum energy exposure deemed to be both effective and safe. This would only be possible if both therapeutic objectives of effectiveness and safety were under the control of the operator, but to a large extent they are. The site intensity at the focal point (the target zone) can be monitored using visual feedback, as evidenced by hyperechoic changes on B- mode US. It is possible to increase energy exposure to obtain these visual changes and to decrease the exposure if the changes become uncontrolled. Our hypothesis is that obtaining visual changes at the focal point can serve as a real-time feed-back to the operator that cytocidal levels of energy are being delivered to the part of the prostate being treated. Implicit in this approach are strong, and we think robust, safety considerations. By controlling the visual change at the threshold level at the focal point, the operator is as certain as possible that the energy is being deposited in the intended area. Moreover, other in-built safety features, such as the reflectivity index in the near field, place an upper boundary on energy absorption in the area abutting rectal mucosa. We termed this approach ‘visually directed’. Using this, the grey-scale changes seen on diagnostic US are actively monitored, and the power adjusted accordingly. For consensus on the types of changes seen, a semiquantitative method of analysis was developed (Appendix), which allows comparison within and between treatments. These ‘Uchida’ changes were named after Toyoaki Uchida (Professor of Urology in Tokai University Hachioji Hospital, Tokyo, Japan) who performed the preliminary clinical work on the Sonablate device.

Using visually directed treatment, the operator aims to generate grey-scale changes throughout the target tissue. During treatment, the power level (energy exposure) is constantly adjusted to achieve Uchida Grade I or II changes (Fig. 3). By obtaining these changes, the operator can control the energy in the target zone that is either on or just below the cavitation threshold. This grey- scale US feedback is also used to provide a ceiling threshold. Grade III changes occur when uncontrolled cavitation occurs in the near field; this is corrected by reducing the energy exposure. Visually directed HIFU therefore takes into account both inter- and intraprostatic differences in acoustic and thermal properties, and allows the user to respond in real-time to the therapy.

Nine men were treated using the algorithm- based protocol (group 1) and 25 men using the visually directed protocol (group 2). All patients were discharged on the day of treatment. Demographic details are given in Table 1; all patients were followed up for =3 months. After therapy, patient status and treatment-related complications were assessed at fixed intervals by visits to the clinic and by telephone consultations with a specialist nurse practitioner. All men were discharged with an indwelling urethral catheter. The PSA level was measured at 3 months after treatment to give a nadir value. Statistical analysis was used to assess the correlation of variables between groups.

Results

Table 1 shows details of the operative variables and results. The difference between the mean PSA nadirs of the groups was significant (P < 0.005). In group 2, 21 of 25 patients achieved PSA nadirs of =0.2 ng/mL at 3 months after treatment; seven patients achieved undetectable PSA values. The mean PSA nadir achieved in group 2 was 0.15 ng/ mL, vs 1.51 ng/mL in group 1.

A trial without catheter was successful at the first attempt in eight of the nine patients in group 1, and 21 of 25 in group 2 (84%). In the 3 months after HIFU, a few patients in each group required flexible cystoscopic investigation. Some also had infective complications, which are listed in Table 1.

Discussion

Visually directed HIFU for organ-confined prostate cancer can produce a low PSA nadir 3 months after the procedure. In the present patients, the mean PSA nadir was significantly lower than that using an algorithm-based protocol for treatment of similar patients, and compares favourably with both brachytherapy and cryotherapy for the treatment of organ-confined prostate cancer15,16. In the Seattle brachytherapy series17 72% of patients with no evidence of disease biochemically achieved PSA nadirs of <0.2 ng/mL, with the mean PSA nadir being 0.25 ng/mL. In the present study we achieved PSA nadirs of =0.2 ng/mL in 84% of patients using the visually directed method, and an undetectable PSA level in just under a third of those treated.

Clinicians familiar with TRUS will acknowledge that the characteristics of prostate glands differ between patients. Even men who have had no previous therapy can have glands of different density and with different patterns of micro- or macro- calcification. Just as the amount of pressure that is required to exert on the scalpel is based upon the real-time characteristics of the tissue it is passing through, so is the amount of energy required to cause ablation within the prostate gland.

We have given the first formal description of grey-scale US changes associated with transrectal HIFU treatment for prostate cancer (Appendix). These ‘Uchida changes’ allow a descriptive analysis of changes seen during therapy and permit a formal system of treatment to be developed, which is consistent between users but flexible according to the gland treated. Grey-scale changes seen on B-mode US have been identified in relation to ablative therapies; these have previously been termed ‘pop-corning’ in relation to HIFU treatment of the prostate, and ‘gas cloud’ formation in relation to radiofrequency ablation in the liver, but have not been quantified for use as a method of real-time feedback.

In the past, cavitation was avoided, as it was assumed to be uncontrollable, and that the risk of cavitation outside the area of interest was too great. Extensive dosimetry studies 7,18 showed that not only are the grey-scale changes visualized on B-mode US associated with histological ablation, but that single pulses of high-intensity ultrasound can produce well circumscribed, predictable volumes of necrosis. It might be argued that, by producing cavitation, the tissue is being ‘over-treated’; in the absence of other real- time methods of detecting thermal ablation, this remains the best method of treatment monitoring. Tissue elastography19 and ultrasound thermometry20 are under development but remain experimental; MRI21 might accurately detect temperature changes, but MRI devices are costly, do not provide feedback as instantaneously as B- mode US, and have not been used clinically in the setting of transrectal prostate HIFU.

Although presently the diagnostic TRUS uses 7 MHz probes and the 4–6 MHz centre frequency band of the Sonablate-500 is not the standard frequency for diagnostic imaging of the prostate, we have had no difficulty in using it for planning and monitoring treatment. This 4–6 MHz frequency band allows excellent visualization of the prostatic margin and grey-scale changes within the gland. Higher frequency TRUS is used in all patients before treatment, and even with the highest ultrasonic resolution the differentiation between benign and malignant prostate is still inaccurate and therefore unnecessary for the purposes of treatment22.

Despite the few patients in each group, the catheter-free rate appears equivalent between them (>80% at the first attempt) with infective complications in ˜10% of patients. This is consistent with other reports using combined prostatic resection and HIFU23. After treatment, most patients have short-term irritative voiding symptoms as a result of the sloughing of prostatic tissue via the urethra. In the visually directed group, more patients underwent flexible cystoscopy. In all cases this was done to investigate irritative and obstructive voiding symptoms, with the result that urethral debris was cleared. The threshold for undertaking a flexible cystoscopy is now considerably higher, as most patients are taught intermittent self-catheterization before treatment, which allows the dislodging of prostatic slough with no need for formal intervention.

We assumed a relationship between the PSA nadir at 3 months and treatment outcome. Data assessing this relationship indicate that this is a justifiable association24, but in that study the outcome was likelihood of disease on prostate biopsy at 6 months after treatment. Although it is logical to assume that this affects the long-term outcome, there are no long-term data to verify it at present; certainly the PSA nadir was shown to correlate with longer term outcome in the context of radical surgery and external beam radiotherapy 25,26.

The present study represents the first reported experience of visually directed HIFU for treating organ-confined prostate cancer. We think that this is the first attempt to standardize the conduct of treatment. Standardization of therapy makes it easier to teach and makes it possible to derive quality standards. Most importantly, standardizing the intervention is the key step in health technology assessment. Once this is done it is possible to start to explore the next phase of investigation, defining the determinants of outcome. This is likely to lead to better case selection and improved conduct of therapy.

Acknowledgements

We are grateful to those at Misonix, Inc. for their ongoing financial support. Rowena Couling (Specialist Nurse Practitioner) for her help with data management and Naren Sanghvi and Focus Surgery for their scientific support.

Conflict of Interest

R. Illing is supported by a grant from Misonix; M. Emberton has acted as a paid consultant to Misonix. Source of funding: Misonix – European distributor of the Sonablate device.

References

  1. Cancer Research UK. UK prostate cancer incidence statistics 2002. Available online. Accessed July 2006
  2. Ahmed S, Lindsey B, Davies J. Emerging minimally invasive techniques for treating localized prostate cancer. BJU Int 2005; 96: 1230–4
  3. Kennedy JE. High-intensity focused ultrasound in the treatment of solid tumours. Nat Rev Cancer 2005; 5: 321–7
  4. ter Haar GR. High intensity focused ultrasound for the treatment of tumors. Echocardiography 2001; 18: 317–22
  5. Dewhirst MW, Viglianti BL, Lora- Michiels M, Hanson M, Hoopes PJ. Basic principles of thermal dosimetry and thermal thresholds for tissue damage from hyperthermia. Int J Hyperthermia 2003; 19: 267–94
  6. Gillett MD, Gettman MT, Zincke H, Blute ML. Tissue ablation technologies for localized prostate cancer. Mayo Clin Proc 2004; 79: 1547–55
  7. Wu F, Wang ZB, Wang ZL et al. Changes in ultrasonic image of tissue damaged by high-intensity ultrasound in vivo. J Acoustic Soc Am 1998; 103: 2869
  8. Sanghvi NT, Fry FJ, Bihrle R et al. Noninvasive surgery of prostate tissue by high-intensity focused ultrasound. IEEE Transactions on Ultrasonics, Ferroelectrics Frequency Control 1996; 43: 1099–110
  9. Fry FJ, Sanghvi NT, Foster RS, Bihrle R, Hennige C. Ultrasound and microbubbles: their generation, detection and potential utilization in tissue and organ therapy – experimental. Ultrasound Med Biol 1995; 21: 1227–37
  10. Madersbacher S, Pedevilla M, Vingers L, Susani M, Marberger M. Effect of high- intensity focused ultrasound on human prostate cancer in vivo. Cancer Res 1995; 55: 3346–51
  11. Foster RS, Bihrle R, Sanghvi N et al. Production of prostatic lesions in canines using transrectally administered high- intensity focused ultrasound. Eur Urol 1993; 23: 330–6
  12. Gelet A, Chapelon JY, Poissonnier L et al. Local recurrence of prostate cancer after external beam radiotherapy: early experience of salvage therapy using high- intensity focused ultrasonography. Urology 2004; 63: 625–9
  13. Uchida T, Ohkusa H, Nagata Y, Hyodo T, Satoh T, Irie A. Treatment of localized prostate cancer using high-intensity focused ultrasound. BJU Int 2006; 97: 56– 61
  14. Gelet A, Chapelon JY, Bouvier R et al. Transrectal high-intensity focused ultrasound: minimally invasive therapy of localized prostate cancer. J Endourol 2000; 14: 519–28
  15. de La Taille A, Benson MC, Bagiella E et al. Cryoablation for clinically localized prostate cancer using an argon-based system: complication rates and biochemical recurrence. BJU Int 2000; 85: 281–6
  16. Storey MR, Landgren RC, Cottone JL et al. Transperineal 125iodine implantation for treatment of clinically localized prostate cancer: 5-year tumor control and morbidity. Int J Radiat Oncol Biol Phys 1999; 43: 565–70
  17. Sylvester JE, Blasko JC, Grimm PD, Meier R, Malmgren JA. Ten-year biochemical relapse-free survival after external beam radiation and brachytherapy for localized prostate cancer: the Seattle experience. Int J Radiat Oncol Biol Phys 2003; 57: 944–52
  18. ter Haar G, Kennedy JE, Wu F, Illing RO. Real-time ultrasound imaging studies of single lesion formation. 4th International Symposium on Therapeutic Ultrasound. Kyoto, Japan 2004, Abstract
  19. Souchon R, Rouviere O, Gelet A et al. Visualisation of HIFU lesions using elastography of the human prostate in vivo: preliminary results. Ultrasound Med Biol 2003; 29: 1007–15
  20. Pernot M, Tanter M, Bercoff J, Waters KR, Fink M. Temperature estimation using ultrasonic spatial compound imaging. IEEE Trans Ultrason Ferroelectr Freq Control 2004; 51: 606–15
  21. Quesson B, de Zwart JA, Moonen CT. Magnetic resonance temperature imaging for guidance of thermotherapy. J Magn Reson Imaging 2000; 12: 525–33
  22. Carey BM. Imaging for prostate cancer. Clin Oncol (R Coll Radiol) 2005; 17: 553–9
  23. Chaussy C, Thuroff S. The status of high-intensity focused ultrasound in the treatment of localized prostate cancer and the impact of a combined resection. Curr Urol Rep 2003; 4: 248–52
  24. Uchida T, Illing RO, Cathcart PJ, Emberton M. To what extent does PSA nadir predict subsequent treatment failure following trans-rectal HIFU for presumed localized adenocarcinoma of the prostate? BJU Int 2006; 98: 537–9
  25. Ray ME, Thames HD, Levey LB et al. PSA nadir predicts biochemical and distant failures after external beam radiotherapy for prostate cancer: a multi-institutional analysis. Int J Radiat Oncol Biol Phys 2006; 64: 1140–50
  26. Shen S, Lepor H, Yaffee R, Taneja SS. Ultrasensitive serum prostate specific antigen nadir accurately predicts the risk of early relapse after radical prostatectomy. J Urol 2005; 173: 777–80
Correspondence

Rowland O. Illing, The Institute of Urology and Nephrology, University College London, London, UK. e-mail: rowland@doctors.org.uk

Abbreviations

HIFU, high-intensity focused ultrasound; US, ultrasonography.

Appendix

Uchida Changes

We devised a method of assessing grey-scale US changes seen during visually directed therapy to allow quantification and comparison in and between treatments. ‘Uchida changes’ were classified as Grades I, II and III depending on whether hyperechoic regions were identified within individual target treatment zones, became confluent between adjacent HIFU treatment exposures, or were seen migrating outside the target treatment zone, respectively. These were then subclassified into ‘a’, ‘b’ and ‘c’ depending upon whether 50% of the focal region was involved in the changes, respectively (Fig. 3). The aim was to see some form of Uchida change every second or third exposure, to confirm that treatment was taking place on or near the cavitation threshold.